The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.
The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.
The Celltrion DiaTrust COVID-19 Ag Rapid Test is intended to detect the presence of SARS-CoV-2, the virus that causes COVID-19, based on a nasopharyngeal (deep inside the nose to the back of the throat) swab sample from patients. The sample is collected by healthcare providers.
Celltrion USA is recalling specific lots of the DiaTrust COVID-19 Ag Rapid Test due to a high number of false positive reports. A false-positive antigen test result may lead to a delay in both the correct diagnosis and treatment for the actual cause of a person’s illness. False-positive results could also lead to more spread of the SARS-CoV-2 virus if presumed positive people are housed together.
Additionally, the tests’ labeling for the affected products includes a shelf life of 18 months. However, the FDA’s emergency use authorization specifies these tests may only be used for 12 months.
The use of the affected product could cause serious adverse health consequences and death.
Although there have been no reports of injuries, adverse health consequences or death associated with the use of the affected products, false positive or false negative results from improper use of these tests could lead to further exposure of uninfected individuals to SARS-CoV-2 virus. There are also serious injury risks if someone who is not trained to collect a nasopharyngeal swab sample attempts to do so.
On February 28, 2022 Celltrion USA sent letters to some affected distributors and customers with the following instructions:
Customers with questions or concerns about this recall may contact Celltrion USA at CelltrionUSA@celltrion.com.
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program using an online form, regular mail, or FAX.
03/17/2022
Sign up to receive email updates on medical device recalls, safety communications, and other safety information.